Keterangan
HD37AB1347 IAQ Monitor is a tool manufactured by Delta Ohm for the analysis of air
quality ( Indoor Air Quality, IAQ) .
The instrument simultaneously measures several parameters: Carbon Dioxide CO2,
Carbon monoxide CO, Temperature, Relative humidity and calculates Dew Point, wet bulb temperature, absolute humidity, mixing ratio, enthalpy and atmospheric
pressure. All this with the P37AB147 SICRAM probe. The probe SICRAM P37B147 does
not measure the Carbon Monoxide CO. Also combined temperature and humidity
SICRAM probes, Hot wire Air speed SICRAM probes, Vane air speed SICRAM probes
and temperature SICRAM probes can be connected to the instrument.
The instrument, with proper procedure, calculates the percentage of outdoor air intake
( % Outside Air) as a function of both carbon dioxide CO2 and temperature and the
Ventilation Rate.
HD37AB1347 data logger has a storage capacity of 67, 600 presets for each of the two
inputs divided into 64 blocks. Use the software DeltaLog10 version 0.1.5.0.
The instrument is equipped with a large dot matrix graphic display with a resolution of
160x160 points. Standards: ASHRAE 62.1-2004, Decree Law 81/ 2008. The rules apply
to all enclosed spaces that may be occupied by people. Should be considered, depending
on air quality, chemical contaminants, physical and biological or outdoor air flow inside inadequately purified ( Ventilation Rate) .
Typical applications of the instrument with the range of sensors mentioned above are:
- Measure IAQ and comfort conditions in schools, offices and indoor.
- Analysis and study of sick building syndrome ( Sick Building Syndrome) and
consequences.
- Verification of HVAC system.
- Investigation of IAQ conditions in factories to optimize the microclimate and improve productivity.
- Audits in Building Automation.
HD37AB1347 Technical specifications
Instrument
Dimensions ( Length x Width x Height) 185x90x40 mm
Weight 470 g ( batteries included)
Materials ABS, rubber
Display Backlit, Dot Matrix
160x160 dots, visible area 52x42 mm
Operating conditions
Operating temperature -5… 50° C
Storage temperature -25… 65° C
Working relative humidity 0 … 85% RH without condensation
Protection degree IP66
Instrument uncertainty ± 1 digit @ 20° C
Power supply
Mains adapter ( code SWD10) 12Vdc/ 1A
Rechargeable batteries 4 1.2V type AA batteries Ni-MH
Autonomy 20 hours with 2200mAh Ni-MH batteries
( with P37AB147 probe connected)
Power absorbed with instrument off < 45¼ A
Security of stored data Unlimited
Connections
Input for probes with SICRAM module T wo 8-pole male DIN45326 connectors
You can connect the following probes to the Indoor Air Quality input: - P37AB147
- P37B147
- Temperature probes equipped with SICRAM
module
- Temperature and Humidity combined probes with SICRAM module
You can connect the following probes to the Temp - Air Velocity input: - Hot-Wire Sensor Air Speed probes with SICRAM
module
- Vane Air Speed probes with SICRAM module
- Temperature probes equipped with SICRAM
module
Serial interface:
Socket: 8-pole M12
Type: RS232C ( EIA/ TIA574) or USB 1.1 or 2.0 not insulated
Baud rate: Between 1200 and 38400 baud. With USB
baud= 460800
Data bits: 8
Parity: N one
Stop bits: 1
Flow control: Xon-Xoff
Cable length: Max 15 m
HD37AB1347
Indoor Air Quality Monitor
227
Air Quality - CO - CO2
Air Quality - CO - CO2
Memory Divided into 64 blocks.
Storage capacity 67600 recordings per each of the 2 inputs.
Logging interval Selectable among: 15, 30 seconds, 1, 2, 5, 10, 15, 20, 30 minutes and 1 hour.